This test concerns the LB (Label Broker) actor. You will use the Order Manager tool to simulate the Label Information Provider actor.
In this test, the Label Broker notify the effective labeled containers production to the Label Information Provider.
So, the SUT (System Under Test), needs to know the labeling instruction, before to label the tubes and to send a notification message to the LIP for the effective labels printing and labeled containers production.
Two steps are necessary in this test :
- First, the LBL Simulator will send the labeling instruction to the SUT (see test "LBL Request Mode For Labeling Instruction")
- Second, the SUT will notify the LBL Simulator for the effective labels printing and labeled containers production.
Instructions
Access the LBL Simulator tool at this location : Order Manager
If it is your first time with this tool, please read the user manual : Order Manager User Manual
Please be remind that if you are logged in your configurations will be private.
- Go to the "Laboratory" menu entry and choose Label Information Provider > Configuration page to begin your test.
- Configure your SUT to send messages to the IP Address and Port displayed on screen.
- Send a message to the LBL Simulator for the effective labels printing and labeled containers production.
- Access the messages exchanged between your SUT and the simulator under Messages > HL7v2 messages.
- Check the simulator has properly received your acknowledgement message.
- The messages exchanged between the simulator and the SUT can be found in the message table on the very same page or from the HL7 Message menu entry. Hit the link on the left side of the raw (first column of the table), to display the Permanent Link to test report for the corresponding HL7 message. (You can access to the hl7v2 report tutorial for more details.) If the validation report status is passed for the message and the response, copy the "test result link" and paste it in Gazelle as the result of this test. For further details, see this tutorial : How enter your pre-connectathon test result in Gazelle (only for the LAW, LCSD and LBL Profiles of the Laboratory Domain which send HL7v2 messages) ?
Evaluation
The validation status must be passed for the two messages of the transaction. The message type must be the right message type, according to IHE.
The Acknowledgment code must be "AA" (MSA-1) in the acknowledgment message.