This test concerns only the Analyzer actor. You will need to communicate with the LAW Simulator (wich is included in the Order Manager Simulator), in order to simulate the LAB-28 transaction of the LAW Profile.
Instructions
Access the LAW Simulator tool at this location : Order Manager
If it is your first time with this tool, please read the user manual : Order Manager User Manual
Please be reminded that if you are logged in your configurations will be private, otherwise it will be public.
In this test, the LAW Simulator will be used to send to the SUT (System Under Test) an AWOS related to a specimen.
As the SUT implements the Analyzer (Analyzer acts as a responder in this test) :
- First at all, check that the selected "Domain for testing" (at the bottom right of the page) is "Laboratory".
- Create or update the configuration corresponding to the SUT.
- Then, go to the "Order Management" menu entry, choose "Analyzer Manager", and click on the "Analytical Work Order Step Broadcast" page to begin the test.
- In the SUT Configurations drop-down list, select the SUT configuration.
-
- For the "Action to perform", select "Create a new order".
- Select the patient to begin the test. Of course, you can create a new patient and a new encounter.(See the Order Manager user manual to have more information).
- Once the patient and the encounter have been selected, go to the "Create specimen" panel and fill all required fields. It is possible to fill some fields randomly in hitting the "Fill with random values" button.
- Hit the "create an order for specimen" button to save the specimen.
- Then, in the "Create order for current specimen" panel, fill all required fields. Once again, it is possible to fill some fields randomly in hitting the "Fill with random values" button.
- For the Result status, choose the value : "Final results: results stored and verified. Can only be changed with a corrected result.".
- At least, hit the "Add to current specimen and send message" button to send the order to the selected SUT.
- Check the simulator has properly received your acknowledgement message.
-
- If you always are in the "Test Report" panel, hit the "Perform another test" button, otherwise return to the "Order Management" menu entry, choose "Analyzer Manager", and click on the "Analytical Work Order Step Broadcast" page.
- For the "Action to perform", select "Cancel an existing order".
- In the "Order Selection" panel, select the Order sent in the first step. You can use the filter option to search a specific Order.(See the Order Manager user manual to have more information).
- A modal panel will ask you to confirm the order cancellation. Say "Yes". The LAW Simulator will send the cancellation order to the SUT.
- Check the simulator has properly received your acknowledgement message.
- First Step : Create and send a new order.
- Second Step : Cancel the AWOS sent in the first step.
The messages exchanged between the simulator and the SUT can be found in the message table on the very same page (see the "Test report" panel) or from the HL7 Message menu entry.
Hit the link on the left side of the raw (first column of the table), to display the Permanent Link to the test report for the corresponding HL7 message. (You can access to the hl7v2 report tutorial for more details.)
If the validation report status is passed for the message and the response, copy the "test result link" and paste it in Gazelle as the result of this test.
For further details, see this tutorial : How enter your pre-connectathon test result in Gazelle (only for the LAW, LCSD and LBL Profiles of the Laboratory Domain which send HL7v2 messages)?
Evaluation
- For all steps :
- The validation status must be passed for the two messages of the transaction. The message type must be the right message type, according to IHE.
- The Acknowledgment code must be "AA" (MSA-1) in the acknowledgment message.
- For the First Step :
- In the ORL^O34^ORL_O34 message, the ORC-1 must be filled with "OK" and ORC-5 with "SC" or "IP".
- For the Second Step :
- In the ORL^O34^ORL_O34 message, the ORC-1 must be filled with "CR" and ORC-5 with "CA".